A Chloride Ion-Selective Potentiometric Sensor Based on a Polymeric Schiff Base Complex

Dyuthi/Manakin Repository

A Chloride Ion-Selective Potentiometric Sensor Based on a Polymeric Schiff Base Complex

Show simple item record

dc.contributor.author Girish Kumar,K
dc.contributor.author Saji, John K
dc.contributor.author Indira, C J
dc.date.accessioned 2010-11-20T06:40:50Z
dc.date.available 2010-11-20T06:40:50Z
dc.date.issued 2005-11-18
dc.identifier.citation Department of Applied Chemistry, Cochin University of Science and Technology en_US
dc.identifier.other Indian Journal of Chemical Technology Vol.13, January 2006, pp. 13-16
dc.identifier.uri http://dyuthi.cusat.ac.in/xmlui/purl/1928
dc.description.abstract This paper describes the fabrication of an ion-selective electrode in which a polymeric Schiff base complex of cobalt(II) is used as the ionophore.The main advantage of the electrode is that it is mechanically stable upto 3 months..The electrode shows a linear response in the range of 2.5 × 10-5-0.5 × 10-1 mol dm-3. The response time of the electrode is 30 s.The pH range at which the electrode works is 3.8 to 6.8. The electrode was found to be selective towards chloride ion in the presence of ions like Na+, Ca2+, Mn2+, ,Fe3+, Co2+, Ni2+, Cu2+, Zn2+, CH3COO-, NO3-, SO42- ,Br- and NO2-. en_US
dc.language.iso en en_US
dc.publisher Indian Journal of Chemical Technology en_US
dc.subject Ion Selective Electodes en_US
dc.subject Chloride Ion en_US
dc.subject Amino Methylated Polystyrene en_US
dc.title A Chloride Ion-Selective Potentiometric Sensor Based on a Polymeric Schiff Base Complex en_US
dc.type Working Paper en_US