dc.contributor.author |
Girish Kumar,K |
|
dc.contributor.author |
Augustine, Pearl |
|
dc.contributor.author |
John, Sareena |
|
dc.contributor.author |
Radecki, Jerzi |
|
dc.contributor.author |
Radecka, Hanna |
|
dc.date.accessioned |
2010-12-03T08:25:45Z |
|
dc.date.available |
2010-12-03T08:25:45Z |
|
dc.date.issued |
2008 |
|
dc.identifier.issn |
0003-2719 |
|
dc.identifier.other |
Analytical Letters, 41: 1144–1157, 2008 |
|
dc.identifier.uri |
http://dyuthi.cusat.ac.in/xmlui/purl/1989 |
|
dc.description.abstract |
The fabrication and analytical applications of two types of potentiometric
sensors for the determination of ketoconazole (KET) are described. The
sensors are based on the use of KET-molybdophosphoric acid (MPA) ion pair
as electroactive material. The fabricated sensors include both polymer membrane
and carbon paste electrodes. Both sensors showed a linear, stable and near
Nernstian slope of 57.8mV=decade and 55.2mV=decade for PVC membrane
and carbon paste sensors respectively over a relatively wide range of KET concentration
(1×10-2-5×10-5 and 1×10-2-1×10-6). The sensors showed a fast
response time of <30 sec and <45 sec. A useful pH range of 3–6 was obtained
for both types of sensors. A detection limit of 2.96 10 5M was obtained for
PVC membrane sensor and 6.91 10 6M was obtained for carbon paste sensor.
The proposed sensors proved to have a good selectivity for KET with respect to a
large number of ions. The proposed sensors were successfully applied for the
determination of KET in pharmaceutical formulations. The results obtained
are in good agreement with the values obtained by the standard method. |
en_US |
dc.description.sponsorship |
Polish Academy of Sciences & Cochin University of Science
and Technology |
en_US |
dc.language.iso |
en |
en_US |
dc.publisher |
Taylor & Francis |
en_US |
dc.subject |
Carbon paste sensor |
en_US |
dc.subject |
Ketoconazole |
en_US |
dc.subject |
Molybdophosphoric acid |
en_US |
dc.subject |
Pharmaceutical formulations |
en_US |
dc.subject |
Polymeric membrane sensor |
en_US |
dc.subject |
Potentiometry |
en_US |
dc.title |
Fabrication of Potentiometric Sensors for the Selective Determination of Ketoconazole |
en_US |
dc.type |
Working Paper |
en_US |
dc.contributor.faculty |
Science |
en_US |
dc.identifier.url |
DOI: 10.1080/00032710802051983 |
en_US |